자료유형 | 학위논문 |
---|---|
서명/저자사항 | Analysis of Metallic Shielding for Reduction of RF Induced Heating of Electrode During MRI for Active Implants. |
개인저자 | Singhal, Krishna. |
단체저자명 | Purdue University. Electrical and Computer Engineering. |
발행사항 | [S.l.]: Purdue University., 2018. |
발행사항 | Ann Arbor: ProQuest Dissertations & Theses, 2018. |
형태사항 | 186 p. |
기본자료 저록 | Dissertation Abstracts International 80-01B(E). Dissertation Abstract International |
ISBN | 9780438369085 |
학위논문주기 | Thesis (Ph.D.)--Purdue University, 2018. |
일반주기 |
Source: Dissertation Abstracts International, Volume: 80-01(E), Section: B.
Adviser: John A. Nyenhuis. |
요약 | The options available to patients with implantable devices are limited. It is because there are multiple interactions between the MRI environment and the implantable medical devices. The three main components of MRI systems- static magnet, RF co |
요약 | The focus of current research is to provide analysis of shielding technique for reduction of RF-induced heating of electrodes during MRI. Shielded leads have been developed as a method to reduce RF-heating responsible for temperature rise at the |
요약 | The information provided by this work is expected to facilitate the development of lead wires with reduced RF-induced heating. The availability of lead wires with reduced heating will allow expanded access to MRI by patients with implantable dev |
일반주제명 | Electrical engineering. Medical imaging. Electromagnetics. |
언어 | 영어 |